Table 2 Experimental results

From: Remote radioactive seed-loading device for permanent brachytherapy of oral cancer with Au-198 grains

Subject

a

b

c

d

e

PA (μSv)

10.2

17.7

6.51

17.2

16.0

PL (μSv)

5.02

4.52

2.99

2.78

6.52

PH (μSv)

5.48

PI (μSv)

10.5

10.0

8.47

8.46

12.0

Total (proposed) (μSv) = PA + PI

20.7

27.7

15.0

25.6

28.0

TL (μSv)

40.3

63.0

17.4

39.7

24.4

TI (μSv)

26.0

Total (traditional) (μSv) = TL + TI

66.3

89.0

43.4

65.7

50.4

Reduction ratio (%)

68.8

68.9

65.4

60.9

44.4

  1. PA exposure dose during the seed-aligning process
  2. PL exposure dose from the applicator in which a seed is loaded and already implanted seeds in the patient during the implantation process
  3. PH exposure dose during the seed implantation to the patient which is estimated from the traditional process
  4. PI exposure dose during the implantation process (PL+PH)
  5. TL exposure dose during the traditional seed-loading process
  6. TI exposure dose from the handling applicator, implanted seeds and prepared applicators beside the doctor during the traditional implantation process